首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Simultaneous determination of trace metals in sewage and sewage effluents by inductively coupled argon plasma atomic emission spectrometry
Authors:MM Moselhy  PN Vijan
Institution:Laboratory Services Branch, Ministry of the Environment, P.O. Box 213, Rexdale, Ontario, M9W 5Ll Canada
Abstract:Inductively coupled plasma excitation sources are adapted for the simultaneous determination of multiple elements in sewage and sewage effluents by atomic emission detection. Digestion with aqua regia in test tubes is recommended for sample preparation. A typical comparison of lead in sewage by atomic emission (x) and atomic absorption (y) gave a least-squares fit of x = 0.98 y + 0.03 with a standard error of 0.5 mg l-1 and a correlation coefficient of 0.996. More complete comparison data by the atomic emission and absorption techniques are presented for other elements.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏