Ion cloud manipulation using the radiofrequency-only-mode as an improvement for high mass detection in fourier transform mass spectrometry
Authors:C B Jacoby  C L Holliman  D L Rempel  M  L Gross
Institution:1. Midwest Center for Mass Spectrometry, Department of Chemistry, University of Nebraska, 68588-0304, Lincoln, NE, USA
Abstract:It has been difficult to achieve the expected high resolving power for high-mass biomolecule ions in Fourier transform mass spectrometry. Our hypothesis is that ion clouds produced by laser desorption or injection are diffuse and produce poor signals. To test the hypothesis, clouds of benzene molecular ions produced by electron ionization were purposefully expanded via magnetron mode excitation and characterized by a new experimental sequence for cloud sectional analysis. The expanded cloud was then successfully focused to the trap center by using a high-pressure dynamic event (radiofrequency-only mode). The expanded cloud in a conventional cubic trap produces no detectable signal, whereas the focused cloud in a compensated trap yields a high-resolution signal with good signal-to-noise ratio.