Selective chemical ionization of nitrogen and sulfur heterocycles in petroleum fractions by ion trap mass spectrometry
Authors:C S Creaser  F Krokos  K E ÓNeill  M J C Smith  P G McDowell
Institution:1. School of Chemical Sciences, University of East Anglia, Norwich, UK
3. BP Research, Sunbury-on-Thames, Middlesex, UK
Abstract:A procedure is reported for the selective ammonia chemical ionization of some nitrogen and sulfur heterocycles in petroleum fractions using ion trap mass spectrometry (ITMS). The ion trap scan routine is designed to optimize the population of ammonium reagent ions and eject from the trap (by radio frequency/direct current isolation) electron ionization products formed during reagent ion formation prior to ionization of the sample. The ITMS procedure is compared with standard ion trap detector and conventional quadrupole ammonia chemical ionization for the determination of nitrogen and sulfur heterocycles in gas oil and kerosine samples. Greatly enhanced selectivity is shown for the ITMS procedure by suppression of competing charge-exchange processes.
Keywords:
本文献已被 ScienceDirect SpringerLink 等数据库收录!