Validation of a Fast and Simple HPLC-UV Method for the Quantification of Adenosine Phosphates in Human Bronchial Epithelial Cells
Authors:Ana Teresa Juarez-Facio  Violaine Martin de Lagarde  Christelle Monteil  Jean-Marie Vaugeois  Ccile Corbiere  Tiphaine Rogez-Florent
Institution:Normandie Univ, UNIROUEN, UNICAEN, ABTE, 76000 Rouen, France; (A.T.J.-F.); (V.M.d.L.); (C.M.); (J.-M.V.); (C.C.)
Abstract:A new HPLC method for the simultaneous quantitative analysis of adenosine triphosphate (ATP), adenosine diphosphate (ADP), and adenosine monophosphate (AMP) was developed and validated. ATP, ADP, and AMP were extracted from human bronchial epithelial cells with a rapid extraction procedure and separated with a C18 column (3 × 150 mm, 2.7 µm) using isocratic elution with a mobile phase consisting of 50 mM of potassium hydrogen phosphate (pH 6.80). The absorbance was monitored at 254 nm. The calibration curves were linear in 0.2 to 10 µM, selective, precise, and accurate. This method allowed us to quantify the nucleotides from two cell models: differentiated NHBE primary cells grown at the air–liquid interface (ALI) and BEAS-2B cell line. Our study highlighted the development of a sensitive, simple, and green analytical method that is faster and less expensive than other existing methods to measure ATP, ADP, and AMP and can be carried out on 2D and 3D cell models.
Keywords:HPLC-UV  BEAS-2B cells  NHBE cells  adenosine phosphates  cellular extraction