首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A Robust Flow-Based System for the Spectrophotometric Determination of Cr(VI) in Recreational Waters
Authors:Tnia C F Ribas  Raquel B R Mesquita  Ana Machado  Joana L A Miranda  Graham Marshall  Adriano Bordalo  Antnio O S S Rangel
Abstract:A flow-based method for the spectrophotometric determination of chromium (VI) in recreational waters with different salinities was developed. Chromium can occur in the environment in different oxidation states with different related physiological properties. With regard to chromium, the speciation is particularly important, as the hexavalent chromium is considered to be carcinogenic. To achieve that purpose, the use of the diphenylcarbazide (DPC) selective colored reaction with the hexavalent chromium was the chosen strategy. The main objective was to develop a direct and simple spectrophotometric method that could cope with the analysis of different types of environmental waters, within different salinity ranges (fresh to marine waters). The potential interference of metal ions, that can usually be present in environmental waters, was assessed and no significant interferences were observed (<10%). For a complete Cr(VI) determination (three replicas) cycle, the corresponding reagents consumption was 75 µg of DPC, 9 mg of ethanol and 54 mg of sulfuric acid. Each cycle takes about 5 min, including the system clean-up. The limit of detection was 6.9 and 12.2 µg L−1 for waters with low and high salt content, respectively. The method was applied for the quantification of chromium (VI) in both fresh and marine water, and the results were in agreement with the reference procedure.
Keywords:sequential injection analysis  GlobalFIA  Cr(VI)  DPC  dynamic water systems
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号