首页 | 本学科首页   官方微博 | 高级检索  
     


Ion Behavior in an Electrically Compensated Ion Cyclotron Resonance Trap
Authors:Brustkern Adam M  Rempel Don L  Gross Michael L
Affiliation:Department of Chemistry, Washington University in St. Louis, One Brookings Drive, Box 1134, St. Louis, Missouri, USA.
Abstract:We recently described a new electrically compensated trap in FT ion cyclotron resonance mass spectrometry and developed a means of tuning traps of this general design. Here, we describe a continuation of that research by comparing the ion transient lifetimes and the resulting mass resolving powers and signal-to-noise (S/N) ratios that are achievable in the compensated vs. uncompensated modes of this trap. Transient lifetimes are ten times longer under the same conditions of pressure, providing improved mass resolving power and S/N ratios. The mass resolving power as a function of m/z is linear (log-log plot) and nearly equal to the theoretical maximum. Importantly, the ion cyclotron frequency as a function of ion number decreases linearly in accord with theory, unlike its behavior in the uncompensated mode. This linearity should lead to better control in mass calibration and increased mass accuracy than achievable in the uncompensated mode.