Evaluation of different calibration strategies for the analysis of pure copper and zinc samples using femtosecond laser ablation ICP-MS
Authors:Heike Traub   Markus W?lle   Joachim Koch   Ulrich Panne   Ralf Matschat   Heinrich Kipphardt  Detlef Günther
Affiliation:(1) BAM Federal Institute for Materials Research and Testing, Richard-Willstaetter-Strasse 11, 12489 Berlin, Germany;(2) ETH Zurich, Laboratory of Inorganic Chemistry, Wolfgang-Pauli-Strasse 10, 8093 Zurich, Switzerland;(3) Department of Chemistry, Humboldt-Universitaet zu Berlin, Brook-Taylor-Strasse 2, 12489 Berlin, Germany
Abstract:Solution-doped metal powder pellets as well as aspirated liquids were used as calibration samples to analyze pure copper and zinc certified reference materials (CRMs) by femtosecond laser ablation ICP-MS. It was demonstrated that calibration by copper pellets resulted in relative deviations up to 20%, whereas fs-LA-ICP-MS among copper-based CRMs led to inaccuracies in the same range unless nominal mass fractions were chosen to be <3 mg/kg. Calibration by zinc pellets generally provided better accuracy. Depending on the analyte considered, deviations below 10% were obtained even for mass fractions close to the limit of quantification. Our data, therefore, indicate solution-doped metal powder pellets to be suitable as calibration samples for fs-LA-ICP-MS of metals. Furthermore, the utilization of liquid standards for calibration was found to result in stronger deviations of up to 50% for both copper and zinc samples which, in addition, turned out to be dependent on the plasma conditions. MediaObjects/216_2009_3061_Figa_HTML.gif
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号