Sensitive voltammetric determination of rutin at an ionic liquid modified carbon paste electrode |
| |
Authors: | Zhang Ya Zheng Jianbin |
| |
Affiliation: | a Institute of Analytical Science/Shaanxi Provincial Key Lab of Electroanalytical Chemistry, Northwest University, Xi’an, Shaanxi 710069, China b Department of Chemistry and Chemical Engineering, Yulin College, Yulin, Shaanxi 719000, China |
| |
Abstract: | An ionic liquid modified carbon paste electrode (IL/CPE) had been fabricated by using hydrophilic ionic liquid 1-amyl-3-methylimidazolium bromide ([AMIM]Br) as a modifier. The IL/CPE was characterized by scanning electron microscope and voltammetry. Electrochemical behavior of rutin at the IL/CPE had been investigated in pH 3.29 Britton-Robinson (B-R) buffer solution by cyclic voltammetry (CV) and square wave voltammetry (SWV). The experimental results suggested that the modified electrode exhibited an electrocatalytic activity toward the redox of rutin. The electron transfer coefficient (α) and the standard rate constant (ks) of rutin at the modified electrode were calculated. Under the selected conditions, the reduction peak current was linearly dependent on the concentration of rutin in the range of 4.0 × 10−8 to 1.0 × 10−5 mol L−1 (r = 0.9998), with a detection limit of 1.0 × 10−8 mol L−1 (S/N = 3). The relative standard deviation (R.S.D.) for six times successful determination of 8.0 × 10−7 mol L−1 rutin was 1.2%. The proposed method was applied to determine rutin in tablet and urine sample. In addition, the IL/CPE exhibited a distinct advantage of simple preparation, surface renewal, good reproducibility and good stability. |
| |
Keywords: | Voltammetry Ionic liquid Carbon paste electrode Rutin |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|