首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A novel biamperometric biosensor for urinary oxalate determination using flow-injection analysis
Authors:Milardovi? Stjepan  Kerekovi? Irena  Nodilo Marijana
Institution:Department of General and Inorganic Chemistry, Faculty of Chemical Engineering and Technology, University of Zagreb, Maruli?ev trg 19, HR-10000 Zagreb, Croatia. stjepan.milardovic@fkit.hr
Abstract:A biosensor for determination of oxalate concentration in urine has been developed by immobilisation of oxalate oxidase and peroxidase on the surface of an interdigitated gold electrode. Enzyme immobilisation was performed using BSA and glutaraldehyde. Biamperometric measurements were made in flow conditions both in aqueous oxalate solutions (tested concentration range between 50 μM and 10 mM) and in real urine samples (tested measuring range between 5 and 100 μM). Optimal working conditions were examined for flow-injection analysis, and good correlation was achieved between added oxalate quantity and the one measured by biosensor in urine matrix (R2 = 0.9983). The influence of some interferences (ascorbic acid, uric acid, paracetamol, acetylsalicylic acid) was also studied using biamperometric measurement mode.
Keywords:Biamperometric  Bienzymatic  Biosensor  Oxalate  Urine
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏