首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Improvement of an enzyme electrode by poly(vinyl alcohol) coating for amperometric measurement of phenol
Authors:Gha-Young Kim
Institution:Department of Environmental Science and Engineering, Gwangju Institute of Science & Technology (GIST), 1 Oryong-dong, Buk-gu, Gwangju 500-712, South Korea
Abstract:A poly(vinyl alcohol) film cross-linked with glutaraldehyde (PVA-GA) was introduced to the surface of a tyrosinase-based carbon paste electrode. The coated PVA-GA film was beneficial in terms of increasing the stability and reproducibility of the enzyme electrode. The electrode showed a sensitive current response to the reduction of the o-quinone, which was the oxidation product of phenol, by the tyrosinase, in the presence of oxygen. The effects of the PVA and PVA-GA coating, the pH, and the GA:PVA ratio on the current response were investigated. The sensitivity of the PVA-GA-Tyr electrode was 130.56 μA/mM (1.8 μA/μM cm2) and the linear range of phenol was 0.5-100 μM. At a higher concentration of phenol (>100 μM), the current response showed the Michaelis-Menten behavior. Using the PVA-GA-Tyr electrode, a two-electrode system was tested as a prototype sensor for portable applications.
Keywords:Tyrosinase  Poly(vinyl alcohol)  Glutaraldehyde  Carbon paste electrode  Phenol
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏