首页 | 本学科首页   官方微博 | 高级检索  
     


Potentiometric measurements in sequential injection analysis lab-on-valve (SIA-LOV) flow-system
Authors:Timo Kikas
Affiliation:a Process Chemistry Centre, c/o Laboratory of Analytical Chemistry, Åbo Akademi University, FIN-20500 Turku-Åbo, Finland
b Institute of Physical Chemistry, Tartu University, Tartu, Estonia
Abstract:Advantages of using sequential injection analysis lab-on-valve (SIA-LOV) in potentiometric measurements are studied with Ca2+ sensitive solid-contact ion-selective electrodes (SC-ISE) and pH electrode based on polyaniline (PANI). Experiments show that Ca2+-SC-ISE requires a stopped flow mode to be used in order to get longer equilibration time. On the other hand, PANI based pH electrode gives better results under flow conditions. SIA-LOV was found to be a flexible solution handling system for potentiometric measurements. The technique gives a possibility to fine-tune the calibration range or even recondition the electrode before every measurement without losing in linear range of calibration.
Keywords:Sequential injection   Lab-on-valve   Ion-selective electrodes
本文献已被 ScienceDirect 等数据库收录!