首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Axial temperature gradient and mobile phase gradient in microcolumn high-performance liquid chromatography
Authors:Hongzhe Tian
Institution:Department of Instrumentation & Analytical Chemistry, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, 457 Zhongshan Road, Dalian 116023, China
Abstract:The effect of axial temperature gradient (ATG) along a microcolumn on the separation performance at both isocratic and gradient elution mode was investigated. A thermostat system was designed to form an ATG along the packed column. Polycyclic aromatic hydrocarbons (PAHs) were separated on a 0.53 mm  × 150 mm i.d. 5 μm C18 microcolumn, with water and acetonitrile as mobile phase. The separation results obtained at mobile phase gradient (MPG) and ATG in microcolumn HPLC were compared with the results performed at ambient conditions. Extrapolated curves of peak width at half height (wh)versus lnk showed that wh is narrower at the same retention time when ATG was applied in addition to MPG. The column efficiency was enhanced 20-30% and the resolution was slightly reduced because of reduction of selectivity at elevated temperature at ATG condition. The RSD of retention time in ATG mode was less than 2.5%.
Keywords:Axial temperature gradient  High temperature liquid chromatography  Micro-HPLC  Polycyclic aromatic hydrocarbons
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号