首页 | 本学科首页   官方微博 | 高级检索  
     


Sterilization of health care products by 5 MeV bremsstrahlung (X ray)
Authors:Yoshishige Sato   Thoru Takahashi   Toshio Saito   Toshio Sato  Masaaki Takehisa
Affiliation:

Radia Industry Co., Ltd., 168 Ooyagi, Takasaki, Gunma, 370, Japan

Abstract:Radiation sensitivities of . pumilus and spores were examined to bremsstrahlung of 5 MeV EB and Cobalt -60 γ rays in order to confirm the effects of radiation and dose rate. Biological indicators (SPORTROL, NAmSA, USA) were irradiated with the X ray in the dose rate range of 4.7–47kGy/h. D-value of spores was 1.4–1.5 kGy, and that of was 1.1–1.3 kGy. The D-values of and have very small dose rate dependences to X ray, and the D-values are similar to those of γ ray. Dose distribution by X-ray and γ irradiation was measured for cartons containing 32 unit dialyzers. The Dmax./Dmin. of the X-ray irradiation (1.2) was smaller than that of γ ray (1.3).
Keywords:Sterilization   D-value   Radiation sensitivity   Dose rate effect   Bremsstrahlung   Converted X ray   Dose uniformity   Dose distribution
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏