首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Effect of delay time and grid voltage changes on the average molecular mass of polydisperse polymers and polymeric blends determined by delayed extraction matrix-assisted laser desorption/ionization time-of-flight mass spectrometry
Authors:Mineo Placido  Vitalini Daniele  Scamporrino Emilio  Bazzano Sebastiano  Alicata Rossana
Institution:Istituto di Chimica e Tecnologia dei Polimeri, CNR Catania, Viale A. Doria 6, 95125 Catania, Italy. gmineo@dipchi.unict.it
Abstract:The dependence of the calculated average molecular mass of a polyethylene glycol with a large polydispersity on the instrumental parameters adopted in the acquisition of mass spectra using delayed extraction matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (DE MALDI-TOFMS) was investigated. It has been shown that a combined effect of delay times and potential gradients can act on the ion cloud in the source chamber affecting both mass resolution and average molecular mass value of the analyzed polymeric sample. Also examined was a blend of two different polymers (a PEG and a PMMA commercial sample having a similar average molecular mass), which presents an additional problem concerning the discrimination among the different polymer species as a function of the experimental conditions. In this work, the best instrumental conditions to obtain both good resolution and a correct average molecular mass for the examined polydisperse sample are reported.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号