首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Quantification in grazing-emission X-ray fluorescence spectrometry
Institution:1. University of Antwerp (UIA), Department of Chemistry, Universiteitsplein 1, B-2610 Antwerp, Belgium;2. Philips Research Laboratories, Prof. Holstlaan 4, 5656 AA Eindhoven, The Netherlands;1. National Research Center Kurchatov Institute, Moscow, Russia;2. University Twente, MESA Inst Nanotechnol, Ind Focus Grp XUV Opt, Enschede, Netherlands
Abstract:In grazing-emission X-ray fluorescence (GEXRF) spectrometry wavelength-dispersive detection can be applied. Much softer radiation and hence lighter elements than in total-reflection X-ray (TXRF) spectrometry can thus be detected. We used simulations to investigate methods of quantification of GEXRF results involving soft characteristic radiation. From these studies, it is concluded that for ultra-thin layers, e.g. the sub-monolayer amounts encountered in semiconductor contamination analysis, calibration plots are linear. For thicker layers, quantification should be performed very carefully because of deviations from linearity due to absorption of radiation and to oscillations in the calibration curve. These oscillations are caused by interference of fluorescence radiation emitted directly towards the detector and radiation reflected at the sample–substrate interface. Suggestions for a judicious choice of measurement conditions are made and the benefits of internal standardisation are discussed.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号