首页 | 本学科首页   官方微博 | 高级检索  
     


The effect of humidity on the accuracy of measurement of an electret radon dosimeter
Authors:R. H. Mahat   D. A. Bradley   Y. M. Amin   C. Y. Wong  L. D. Su
Affiliation:

a Department of Physics, Applied Radiation Res Laboratory, University of Malaya, 50603 Kuala Lumpur, Malaysia

b School of Physics, University of Exeter, Exeter EX4 4QL, UK

Abstract:Test has been conducted for the response of electret radon dosimeters with varying levels of humidity. At a constant value of radon concentration, measured values have been found to increase linearly with increasing humidity, from 30% RH up to a value of 85% RH. On this basis we argue for the establishment of a humidity-dependent calibration factor.
Keywords:Electret   Humidity   Radon
本文献已被 ScienceDirect 等数据库收录!