An enzyme reactor electrode for determination of amino acids |
| |
Authors: | Gillis Johansson Kerstin Edström Lars ögren |
| |
Affiliation: | Department of Analytical Chemistry, University of Umeå, S-90187 UmeåSweden |
| |
Abstract: | L-Leucine can be determined with an enzyme reactor electrode containing L-amino acid oxidase immobilized with glutaraldehyde to glass. The reactor also contains immobilized catalase which splits the hydrogen peroxide formed. Oxygen for the reaction is also supplied by adding hydrogen peroxide to the samples. The electrode is an ammonia gas sensor. The calibration curve is strictly linear with Nernstian slope between 3·10-5 and 10-3 M leucine. |
| |