The flameless atomic absorption spectrometric determination of aluminum with a carbon atomization system |
| |
Authors: | Toshihisa Maruta Keiichi Minegishi Giichi Sudoh |
| |
Affiliation: | Central Laboratory of Research and Development, Chichibu Cement Co., Ltd., Ohmiya, Chichibu-shi, Saitama-ken Japan |
| |
Abstract: | The atomic absorption spectrometry of aluminum with a carbon tube atomizer is described, with particular reference to the heating cycle, argon flow rate and amplifier time constant. Interferences are eliminated by pre-atomization heating or with a hydrogen-argon-entrained air flame. Under optimal conditions, the sensitivity (1 % absorbance) is 5·10?11g, and the relative standard deviation is 2.3 for 2.5 ng Al. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|