Successive high-precision determination of calcium and magnesium in sea water with a new probe photometer |
| |
Authors: | T Anfält A Granéli |
| |
Institution: | Department of Analytical Chemistry, University of Gothenburg, Fack, S-402 20 Göteborg 5 Sweden |
| |
Abstract: | A high-precision method for determining calcium and magnesium successively in sea water is described. The calcium end-point is evaluated with zinc—zincon as indicator, while magnesium is determined with calcon. The computerized method utilizes a new probe photometer based on modern optoelectronic components. The precision for calcium and magnesium is 0.3 ‰ and 0.4 ‰, respectively. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|