首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of copper(II) ion-selective electrodes for measurements at micromolar concentrations
Authors:Derek Midgley
Affiliation:Central Electricity Research Laboratories, Kelvin Avenue, Leatherhead, Surrey KT22 7SE England
Abstract:Four types of copper ion-selective electrodes have been tested for determining copper at concentrations below lO-6 mol 1l-1. None of the electrodes has a Nemstian response in dilute copper solutions in this concentration range, though their responses are linear in pCu buffer solutions. The causes of the deviations are a direct redox effect in the case of an electrode with a Cu1.8,Se single crystal membrane, production of copper ions by oxidation of the membrane itself in Ag2S—CuS membrane electrodes, and a combination of the two in the case of the R??i?ka Selectrode. The electrode potentials are affected by the oxygen content and pH of the sample solution and the condition of the membrane surface. Precision tests on two types of electrode are described.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏