Fabrication and testing of the recoil mass spectrometer at Bombay Pelletron
Authors:S Nagaraj  H C Jain  P K Joshi  S D Paul  R Palit  H V Panchal  B S Naidu  A Chatterjee  A Navin
Institution:(1) Tata Institute of Fundamental Research, 400 005 Mumbai, India;(2) Bhabha Atomic Research Centre, 400 085 Mumbai, India
Abstract:A recoil mass spectrometer (RMS) has been designed, fabricated and installed at the 15° S beam-line of the Pelletron at TIFR. The RMS consists of a quadrupole doublet just after the target chamber followed by an ‘electrostatic deflector’, a magnetic dipole and a second electrostatic deflector. The recoils produced in the 12C + 58Ni reaction using 60 MeV 12C beam were focussed with the help of electric and magnetic fields and detected in a strip detector placed at the focal plane of the RMS. Further testing of the spectrometer to obtain mass resolution and efficiency are in progress.
Keywords:Charged particle spectrometers  electric and magnetic  fusion reaction  evaporation residues