Abstract: | We present a miniaturized Fourier transform spectrometer (FTS) based on optical microelectromechanical system technology. The FTS is a Michelson interferometer with one scanning mirror. A new type of electrostatic comb drive actuator moves the mirror. We have measured a nonlinearity of the driving system of +/-0.5 mum for a displacement of 38.5 mum . A method is presented to correct the spectrum to get rid of the nonlinearity. The driving reproducibility is +/-25 nm. The measured resolution of the spectrometer after the phase correction is 6 nm at a wavelength of 633 nm. |