首页 | 本学科首页   官方微博 | 高级检索  
     


Tetracycline selective electrode based on molecularly imprinted polymer particles
Authors:Zhi Yong Guo  Pan Pan Gai  Jing Duan  Hui Na Zhang  Sui Wang
Affiliation:Faculty of Materials Science and Chemical Engineering, The State Key Laboratory Base of Novel Functional Materials and Preparation Science, Ningbo University, Ningbo 315211, China
Abstract:Tetracycline selective electrode using molecularly imprinted polymer particles as quasi-ionophore was constructed the first time, and its performance was carefully characterized. Due to the specific recognition of tetracycline by the particles, the selectivity coefficients for routine interferences were less than 10−4. Benefited from the absence of tetracycline in the sensitive membrane and the optimized composition of the inner filling solution, the limit of detection of the electrode was reduced to about 2.5 × 10−8 mol/L. It exhibited a good electrode slope 57.6 mV/decade near the theoretical Nernstian one, with a wide linear working range from 6.0 × 10−8 to 1.0 × 10−3 mol/L. The fabricated electrode should be used in pH 2−4, response time of which was less than 200 sec when the concentration of tetracycline was higher than 1.0 × 10−6 mol/L and no more than 30 min at the concentration of 1.0 × 10−8 mol/L.
Keywords: Ion selective electrode  Molecular imprinting polymer  Quasi-ionophore  Tetracycline
本文献已被 CNKI 万方数据 ScienceDirect 等数据库收录!
点击此处可从《中国化学快报》浏览原始摘要信息
点击此处可从《中国化学快报》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号