Performance of open tubular columns as a function of tube diameter and liquid phase film thickness |
| |
Authors: | L. S. Ettre |
| |
Affiliation: | (1) Chromatography Division, The Perkin-Elmer Corporation, 08656 Norwalk, CT, U.S.A. |
| |
Abstract: | Summary Wall-coated, open-tubular (capillary) columns prepared from different diameter tubing, with different liquid phase film thickness, are compared with each other and with packed and support-coated open-tubular columns. The comparison is based on the variation of the phase ratio and the capacity factor, and includes column efficiency (HETP, theoretical plate number), resolution, retention time, and sample capacity. Problems associated with the evaluation of the sample capacity are outlined. The influence of temperature on column performance is discussed in detail. Finally, the possibilities of short, wide-bore open-tubular columns prepared with a thick liquid-phase film are demonstrated.Parts of this paper were presented at the 35th Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, Atlantic City NJ, March 5–9, 1984, and at the 20th International Symposium on Advances in Chromatography, New York NY, April 16–18, 1984. |
| |
Keywords: | Gas chromatography Open tubular (capillary) columns Sample capacity Column comparison Film thickness Column diameter |