Application of calorimeters for 5 MeV EB and bremsstrahlung dosimetry
Authors:Toshio Sato   Toru Takahashi   Toshio Saito  Masaaki Takehisa

Arne Miller

Affiliation:

Radia Industry Co., Ltd., 168 Ooyagi, Takasaki, Gunma, 370, Japan

Riso National Laboratory, DK 4000, Roskilde, Denmark

Abstract:Graphite and water calorimeters, which were developed for use a 10 MeV electron beams (EB) at Riso National Laboratory, were used for process validation and routine dosimeter calibration at a 5 MeV EB. Water calorimeters were used for reference measurements for 5 MeV EB, the response was found to be directly proportional to the beam current and the variation among three water calorimeters was less than ± 2 % in the range of 10 to 40 kGy. CTA, PMMA, RCD dosimeters were calibrated by irradiating the dosimeters and water calorimeters simultaneously. The water calorimeters was proved to be an useful tool at 5 MeV EB. Graphite calorimeters gave reproducible readings within 3.3 % relative errors (95% confidence level) for X-ray measurement.