Abstract:A new quantitative method has been developed for real-time mapping of temperature changes induced by high intensity focused ultrasound (HIFU). It is based on the temperature dependence of the T1 relaxation time and the equilibrium magnetization. To calibrate the temperature measurement, the functional relationship between T1 and temperature was examined in different samples of porcine muscle and fatty tissue. The method was validated by a comparison of calculated temperature maps with fiber-optic measurements in heated muscle tissue. The experiment showed that the accuracy of the MR method for temperature measurements is better than 1 degree C. Since the acquisition time of the employed MR sequence takes only 3 s per slice and the calculation of the temperature map can be performed within seconds, the imaging technique works nearly in real-time. The temperature measurement could be realized during HIFU showing no disturbances by ultrasound sonication. In comparison to other MR approaches, the advantages of the introduced method lie in a sufficient accuracy and time resolution combined with a reasonable robustness against motion as well as the feasibility for temperature monitoring in fatty tissues.