A method to reduce the equilibration time prior to data capture in ampoule-based isothermal microcalorimetric studies
Authors:Koji Urakami  Anthony E. Beezer  
Affiliation:

a Medway Sciences, NRI University of Greenwich, Medway University Campus, Chatham Maritime, Kent ME4 4TB, UK

b Analysis and Quality Control Department, Takeda Chemical Industries, Ltd., 2-17-85 Jusohonmachi Yodogawa-ku, Osaka 532-8686, Japan

Abstract:This paper reports a technique for heat conduction isothermal microcalorimeters, which allows, by means of a lowering apparatus, the sample and reference ampoules to be lowered very slowly from the pre-equilibration position to the measuring position. The purpose is to minimize the effect on (total) equilibration time arising from both the pre-equilibration time and the dissipation time (i.e. the time to dissipate the thermal shock which occurs when the vessels are lowered manually). Measurements using 3 ml glass ampoules filled with water, as sample and reference, at 25 and 60 °C, showed that the extent of thermal shock was drastically diminished as the lowering speed decreased, and that the associated heat quantity was reduced to less than 0.2 mJ when lowering occurred over a period of more than 180 s. The dissipation time, the time-period required to dissipate the thermal shock, was also shortened to less than 10 min; the standard manual lowering procedure dissipation time being ca. 25 min. In practice experimental measurements of the imidazole catalysed hydrolysis of triacetin and of the solid state oxidation of ascorbic acid showed that the dissipation time was not shortened when the initial power observed was more than 5 μW, however, it was significantly shortened when the initial power observed was around ≤1 μW. The proposed ampoule lowering procedure could be expected to bring about a saving in the total measurement time for reactions with low initial power, such as those associated with long-term stability studies of relatively stable pharmaceuticals. The described procedure also eliminates operator-induced effects associated with manual lowering of ampoules. In principle the device described also would permit automated loading protocols to be developed.
Keywords:Isothermal microcalorimeter   Ampoule   Thermal shock   Long-term stability
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号