首页 | 本学科首页   官方微博 | 高级检索  
     


Suitability of laser Doppler velocimetry for the calibration of pressure microphones
Authors:A. Degroot  B. Gazengel
Affiliation:a Laboratoire d’Acoustique de l’Université du Maine - UMR CNRS 6613 -, Av. O. Messiaen 72085 Le Mans, cedex 9, France
b Acoustics and Fluid dynamics Group, School of Physics, University of Edinburgh Kings Buildings, Edinburgh, UK
Abstract:The details of a new approach for absolute calibration of microphones, based on the direct measurement of acoustic particle velocity using laser Doppler velocimetry (LDV), are presented and discussed. The calibration technique is carried out inside a tube in which plane waves propagate and closed by a rigid termination. The method developed proposes to estimate the acoustic pressure with two velocity measurements and a physical model. Minimum theoretical uncertainties on the estimated pressure and minimum measurable pressure are calculated from the Cramer Rao bounds on the estimated acoustic velocity amplitude and phase. These uncertainties and the minimum measurable pressure help to optimize the experimental set up. Acoustic pressure estimations performed with LDV are compared with acoustic pressures obtained with a reference microphone. Measurements lead to a minimum bias of 0.006 dB and a minimum uncertainty of 0.013 dB on the acoustic pressure estimation for frequencies 1360 Hz and 680 Hz.
Keywords:43.38.Kb   43.58.Vb   43.60.Qv