Quality assurance and quality control for thermal/optical analysis of aerosol samples for organic and elemental carbon
Authors:Judith C. Chow  John G. Watson  Jerome Robles  Xiaoliang Wang  L.-W. Antony Chen  Dana L. Trimble  Steven D. Kohl  Richard J. Tropp  Kochy K. Fung
Affiliation:(1) Desert Research Institute, Nevada System of Higher Education, 2215 Raggio Parkway, Reno, NV 89512, USA;(2) Institute of Earth Environment, Chinese Academy of Sciences, 10 Fenghui South Road, Xi’an High-Tech Zone, Xi’an, 710075, China;(3) Atmoslytic, Inc., 24801 Alexandra Court, Calabasas, CA 91302, USA
Abstract:Accurate, precise, and valid organic and elemental carbon (OC and EC, respectively) measurements require more effort than the routine analysis of ambient aerosol and source samples. This paper documents the quality assurance (QA) and quality control (QC) procedures that should be implemented to ensure consistency of OC and EC measurements. Prior to field sampling, the appropriate filter substrate must be selected and tested for sampling effectiveness. Unexposed filters are pre-fired to remove contaminants and acceptance tested. After sampling, filters must be stored in the laboratory in clean, labeled containers under refrigeration (<4 °C) to minimize loss of semi-volatile OC. QA activities include participation in laboratory accreditation programs, external system audits, and interlaboratory comparisons. For thermal/optical carbon analyses, periodic QC tests include calibration of the flame ionization detector with different types of carbon standards, thermogram inspection, replicate analyses, quantification of trace oxygen concentrations (<100 ppmv) in the helium atmosphere, and calibration of the sample temperature sensor. These established QA/QC procedures are applicable to aerosol sampling and analysis for carbon and other chemical components.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号