首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of an accurate calibration and spectral standardization procedure for Raman spectroscopy
Authors:Hutsebaut Didier  Vandenabeele Peter  Moens Luc
Institution:Department of Analytical Chemistry, Ghent University, Proeftuinstraat 86, B-9000 Ghent, Belgium. didier.hutsebaut@Ugent.be
Abstract:Due to modern developments Raman spectroscopy has evolved into a fast vibrational technique. Detailed fingerprints in combination with non-destructivity and minimal sample preparation has allowed the construction of reference libraries in a variety of research fields. Long-term stability and comparability are important characteristics when developing reference libraries. In addition, small shifts in highly similar spectra of different samples may limit the full potential of Raman spectroscopy. Since libraries often contain a large number of different and/or highly similar spectra, it is important that each data point in all the spectra corresponds to the exact Raman wavenumber. This is often not the case, due to shifts in optical pathway and/or shifts in laser wavelength. This paper describes a complete calibration protocol (wavelength and intensity) and evaluates the procedure for both short and long term stability, by means of 60 randomly selected measurement sessions spread over a period of nine months. A two-step standardization procedure is proposed to deal with spectral shifts.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏