首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A gas‐filled electroluminescence detector for EDXRF Spectrometry
Abstract:The design and features of a sealed‐off proportional scintillation gas‐filled detector (GSPC) are described. The use of a 50 µm Be window and an Xe‐based gas mixture provides high efficiency of registration in the range 1.5–25 keV. The energy resolution changes from 18% for Al Kα (1.49 keV) to 4.2% for Ag Kα (22.2 keV). When the count rate on the unit output is 5 × 104 s?1, the energy resolution at the Mn Kα line does not exceed 9% at an entrance window area near 3 cm 2 . The detector is operated at room temperature and does not require maintenance during operation. The service life of the detector is over 5 years and the radiation lifetime is practically unlimited. The detector is designed for use in EDXRF spectrometers instead of gas‐filled proportional counters. Copyright © 2005 John Wiley & Sons, Ltd.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号