Tube-probe atomisation in electrothermal atomic absorption spectrometry
Affiliation:1. Tampere University of Technology, Tampere, Finland;2. Capgemini, Düsseldorf, Germany;3. University of Groningen, Groningen, The Netherlands;4. Utah Valley University, Orem, UT, USA
Abstract:A method is described in which liquid samples are deposited by a conventional autosampler on to a tubular graphite probe positioned inside a 7.5 mm i.d. Pye Unicam SP-9 atomiser cuvette. The tube-probe has certain advantages in comparison with the flat probe systems described previously. In particular, the precision of determinations in acid media is improved since the sample solution is better confined within the atomiser and hence the effects of droplet spreading are curtailed. Also, the tube-probe is shown to reduce diffusional loss effects at high atomiser temperatures, in comparison with flat probe operation. The characteristic mass values obtained for volatile and medium volatile elements are similar for tube and flat probes. Since a greater sample volume (up to 40 μl) can be deposited and dried in the tube-probe, improved detection limits are anticipated for these elements. For more involatile elements, the greater mass of the tube-probe results in poorer sensitivity by a factor of × 3 for V and × 4 for Cr.