首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of microquantities of nitrogen in compounds containing fluorine by an automatic rapid nitrogen analyser
Authors:Judit Hegedüs
Affiliation:(1) Research Institute for Pharmaceutical Chemistry, Budapest, Hungary
Abstract:Summary It was observed that when the Heraeus automatic nitrogen analyser is used for organic fluorine compounds, the fluorine (particularly as the trifluoromethyl group) interferes. The cause and extent of the interference were established by mass spectroscopy; the cause is the formation of trifluoromethane and polytetrafluoroethylene. The interference was found to be eliminated by addition of potassium perchlorate as additive to the sample boat.
Zusammenfassung Bei Verwendung des automatischen Geräts zur Stickstoffbestimmung von Heraeus zur Analyse fluorhältiger organischer Verbindungen stört das Fluor (besonders wenn es als Trifluormethylgruppe vorliegt). Ursache und Ausmaß dieser Störung wurden massenspektroskopisch untersucht. Die Ursache ist die Bildung von Trifluormethan und Tetrafluoräthan. Die Störung läßt sich durch Zusatz von Kaliumperchlorat in das Einwaageschiffchen beseitigen.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号