首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Processing of bone samples for the determination of ultra low-levels of uranium and plutonium
Authors:D W Efurd  R E Steiner  S P LaMont  J A Musgrave  D L Kottmann
Institution:(1) Los Alamos National Laboratory, P. O. Box 1663, MS J514, Los Alamos, NM 87545, USA;(2) Los Alamos National Laboratory, P. O. Box 1663, MS J514, Los Alamos, NM 87545, USA;(3) Los Alamos National Laboratory, P. O. Box 1663, MS J514, Los Alamos, NM 87545, USA;(4) Los Alamos National Laboratory, P. O. Box 1663, MS J514, Los Alamos, NM 87545, USA;(5) Los Alamos National Laboratory, P. O. Box 1663, MS J514, Los Alamos, NM 87545, USA
Abstract:Summary We have developed cleanroom compatible techniques for processing bone samples for characterization of their uranium and plutonium content. The bone samples are dried and ashed in quartz crucibles placed inside cleanroom compatible thermal ashing furnaces. The bone ash is dissolved in ultra-pure acids prepared by sub-boiling distillation. The uranium and plutonium in the samples are isolated and purified by ion-exchange chromatography and measured by thermal ionization mass spectrometry. The technique is capable of detecting 74 picograms of 238U and 8 femtograms of 239Pu in 100 mg bone ash samples. If the ash contains larger amounts of uranium and plutonium, the technique can be used to isotopically fingerprint the material to identify potential origins.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号