首页 | 本学科首页   官方微博 | 高级检索  
     


Silver- and Coomassie-staining protocols: detection limits and compatibility with ESI MS
Authors:Winkler Christiane  Denker Katrin  Wortelkamp Stefanie  Sickmann Albert
Affiliation:Rudolf-Virchow-Center, DFG-Research Center for Experimental Biomedicine, University of Würzburg, Würzburg, Germany.
Abstract:Staining protocols for PAGE have to be sensitive and should not impair further MS analysis of selected samples. In this study, the MS compatibility of different silver- and Coomassie-staining protocols with a nano-LC-MS/MS system was systematically elucidated. Altogether, 13 different silver-staining, 1 imidazole-staining and 2 Coomassie-staining protocols were used and compared to each other for their achieved sequence coverage and their detection sensitivity. Three proteins were used as model proteins (bovine serum albumin, rabbit L-lactate dehydrogenase, bovine milk beta-lactoglobulin) in decreasing concentration (12 pmol down to 30 fmol) and different staining protocols were applied. The conclusion of this study is that two silver-staining protocols (Blum, H. et al.,. Electrophoresis 1987, 8, 93-99 and Shevchenko, A. et al.,. Anal. Chem. 1996, 68, 850-858) combine good sequence coverage and good sensitivity and are recommended for nano-LC-MS/MS analysis.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号