首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A PDMS sheath flow cuvette for high‐sensitivity LIF measurements in CE
Authors:Cameron D Skinner
Institution:Department of Chemistry and Biochemistry, Concordia University, Montréal, Quebec, Canada
Abstract:A simple method for producing a sheath flow cuvette in PDMS suitable for post‐column detection in CE is described. Two types of cuvette were investigated. In the first, the sheath flow channel had a round cross‐section of approximately 635 μm diameter, whereas the second cuvette had a 300×300 μm2 square channel. Both cuvettes produced laminar flows that ensheathed the separation capillary's effluent allowing sensitive fluorescence measurements. The elasticity of the PDMS allowed the 300×300 μm2 square sheath flow channel to expand uniformly and accommodate the larger 330–340 μm od round separation capillary, producing a self‐aligning cuvette with robust mechanical properties. With this cuvette, linear calibrations of over five orders of magnitude and 15–30 zmol fluorescein detection limits were obtained for 12 and 50 μm id capillaries.
Keywords:CE  LIF  Sheath flow cuvette
设为首页 | 免责声明 | 关于勤云 | 加入收藏