A Simple and Sensitive Method for the Amperometric Detection of Trace Chromium(VI) Based on Prussian Blue Modified Glassy Carbon Electrode
Authors:Sujie Xing  He Xu  Guoyue Shi  Junshui Chen  Liping Zeng  Litong Jin
Institution:1. Department of Chemistry, East China Normal University, Shanghai 200062, P.?R. China;2. Shanghai Entry‐Exit Inspection and Quarantine Bureau of P.?R.C., Shanghai 200135, P.?R. China
Abstract:A simple and sensitive method for the amperometric detection of trace amount of chromium(VI) using a Prussian blue modified glassy carbon electrode (PB/GCE) is described here. The Everitts salt/Prussian blue redox couple of the PB film was found to mediate the Cr(VI) reduction, and the mechanism of electrochemical reaction was investigated. The effects of PB film thickness, applied potential and electrolyte solution on the current response of Cr(VI) reduction were thoroughly studied. Under the optimized conditions, the PB/GCE provided a wide linear range for Cr(VI) analysis from 0.5 to 200 ppb with excellent sensitivity (15±0.2 nA/ppb) and low detection limit (0.15 ppb). In addition, the modified electrode showed excellent stability, reproducibility and good resistance to other metal ions and surfactants. Finally, the proposed method was applied to detect trace Cr(VI) in wastewater with satisfactory results. The great advantages of the method were characterized by the simplicity, ease of preparation, stability, short analysis time and cost‐efficiency.
Keywords:Chromium  Prussian blue  Amperometric detection  Heavy metals  Trace analysis