首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Double Modification of Electrode Surface for the Selective Detection of Epinephrine and Its Application to Flow Injection Amperometric Analysis
Authors:Guang‐Ri Xu  Xin‐hua Qi  Fengxia Yang  Jae‐Joon Lee  Ming‐Lu Xu  Yu‐Ping Zhang  Sunghyun Kim
Institution:1. Henan Institute of Science and Technology, Xinxiang 453003, P.?R. China;2. Department of Applied Chemistry, Konkuk University, Chungju 380‐701, Korea;3. Department of Bioscience and Biotechnology, Konkuk University, Seoul 143‐701, Korea
Abstract:A glassy carbon electrode having two polymer layers has been applied to selectively detect epinephrine. The inner layer formed by electropolymerization of macrocyclic nickel complex functioned as an electrocatalyst for epinephrine oxidation and the outer layer composed of hydrolyzed polyurethane γ‐benzyl L ‐glutamate as a screening layer. Differential pulse voltammetry showed almost 100% recovery of epinephrine even in 100‐fold excess of interferents. When applied to a dual glassy carbon electrode as an amperometric detector in flow injection analysis, a linear response over 0.1 μM and 10 μM was obtained. Recovery tested for 5‐fold diluted human urine samples was 97.5%.
Keywords:Double modification  Epinephrine  Ni(II) macrocyclic complex  Differential pulse voltammetry  Flow injection analysis  Medicinal chemistry