首页 | 本学科首页   官方微博 | 高级检索  
     


Ex Situ Prepared Antimony Film Electrode for Electrochemical Stripping Measurement of Heavy Metal Ions
Authors:Vasko Jovanovski  Samo B. Hočevar  Bozidar Ogorevc
Affiliation:Analytical Chemistry Laboratory, National Institute of Chemistry, Hajdrihova 19, SI‐1000 Ljubljana, Slovenia
Abstract:The antimony film electrode (SbFE) was prepared ex situ for anodic and adsorptive stripping voltammetric measurement of selected heavy metal ions. The electrode revealed good linearity for Cd(II) and Pb(II) in a nondeaerated solution of 0.01 M HCl in the examined concentration range from 25 to 80 μg L?1 with limits of detection of 1.1 μg L?1 for Cd(II) and 0.3 μg L?1 for Pb(II) and an excellent reproducibility. The preplated SbFE was also preliminary tested for measuring low levels of Ni(II) using adsorptive stripping voltammetry exhibiting good linearity and sensitivity in combination with only a 30 s deposition step.
Keywords:Antimony film electrode  Anodic stripping voltammetry  Adsorptive stripping voltammetry  Cadmium  Lead  Nickel