首页 | 本学科首页   官方微博 | 高级检索  
     


Improved sensitivity and mass range in time-of-flight bioaerosol mass spectrometry using an electrostatic ion guide
Authors:Gregg?A.?Czerwieniec,Scott?C.?Russell,Carlito?B.?Lebrilla  author-information"  >  author-information__contact u-icon-before"  >  mailto:cblebrilla@ucdavis.edu"   title="  cblebrilla@ucdavis.edu"   itemprop="  email"   data-track="  click"   data-track-action="  Email author"   data-track-label="  "  >Email author,Keith?R.?Coffee,Vincent?Riot,Paul?T.?Steele,Matthias?Frank,Eric?E.?Gard
Affiliation:Department of Chemistry, University of California-Davis, One Shields Avenue, Davis, CA 95616, USA.
Abstract:Bioearosol mass spectrometry (BAMS) analyzes single particles in real time from ambient air, placing strict demands on instrument sensitivity. Modeling of the BAMS reflectron time of flight (TOF) with SIMION revealed design limitations associated with ion transmission and instrument sensitivity at higher masses. Design and implementation of a BAMS linear TOF with electrostatic ion guide and delayed extraction capabilities has greatly increased the sensitivity and mass range relative to the reflectron design. Initial experimental assessment of the new instrument design revealed improved sensitivity at high masses as illustrated when using standard particles of cytochrome C (m/z approximately 12,000), from which the compound's monomer, dimer (m/z approximately 24,000) and trimer (m/z approximately 36,000) were readily detected.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏