首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Disposable Biamperometric Capillary-Fill Device for Glucose
Authors:Chengxiao Zhang  Jian Li
Institution:(1) Department of Chemistry, School of Chemistry and Materials Science, Shaanxi Normal University, Xirsquoan 710062, P.R. China, CN
Abstract:A novel disposable biamperometric capillary-fill device for glucose was fabricated using thick-film technology and incorporating a polymer mediator and glucose oxidase. A simple electrode system consisting of two identical carbon screen-printed electrodes was used as a transducer. The reagent ink of glucose oxidase and Os(bpy)2(PVP)10Cl]Cl was drop-coated on the biosensor substrate between the strips of the electrode. A microliter biamperometric capillary-fill device was constructed by adhesion of a glass cover and the biosensor substrate. A drop of test solution was automatically introduced into the microchannel by capillary action, and then a water-soluble layer containing glucose oxidase was solubilized and quickly dispersed throughout the volume of the microchannel. With a voltage of +200thinspmV applied, the current response was directly proportional to the concentration of glucose in the solution. The disposable device for the analysis of glucose shows a linear response range from 0 to 15thinspmM and is virtually insusceptible to interfering species such as 1thinspmM ascorbic acid. The device designed was successfully applied to the determination of glucose in human serum.
Keywords:: Biosensor  glucose  disposable electrode  biamperometry  osmium polymer  
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏