Abstract: | ABSTRACT Assay procedures based on UV spectrophotometry and high-performance liquid chromatography (HPLC) have been developed for the determination of meloxicam in tablet formulations. The HPLC method used a reversed-phase C18 column with 0.05M Tris acetic acid buffer - tetrabutylammonium reagent–acetonitrile as eluent, and UV detection at 360nm with isoxicam as the internal standard. The UV method was based on measuring an acidic solution of the drug at 341nm. A comparison was established in terms of linearity, sensitivity, precision, and accuracy. Both methods were simple and rapid. HPLC was more precise and more accurate, the UV technique was slightly faster. |