首页 | 本学科首页   官方微博 | 高级检索  
     


Lead(II)-Selective Membrane Electrode Based on Tetraphenylporphyrin
Abstract:ABSTRACT

A PVC membrane electrode for Pb2 ions based on tetraphenylporphyrin was prepared. The sensor exhibits a Nernstian response for lead ions over a wide concentration range (1.0 x 10?5-1.0 x 10?2 M). The limit of detection is 8.5 x 10?6M. It has a response time of 15 s and can be used for at least three months without any divergence in potential. The proposed electrode shows a fairly good discriminating ability towards Pb2 ion in comparison to some alkali, alkaline earth, transition and heavy metal ions. The electrode can be used in the pH range 5.0 to 7.5. It was used as an indicator electrode in potentiometric titration of lead ion.
Keywords:Lead ion-selective electrode  PVC membrane  tetraphenylporphyrin  potentiometry
设为首页 | 免责声明 | 关于勤云 | 加入收藏