首页 | 本学科首页   官方微博 | 高级检索  
     


Ion Chromatographic Determination of Traces of Sodium,Magnesium and Chlorine in Gadolinium Nitrate
Abstract:Abstract

Two independent and sensitive ion chromatographic methods with suppressed conductivity were developed for determination of traces of Cl, Na, and Mg in gadolinium-nitrate. Na-Mg was determined by cation-exchange column after matrix separation, whereas Cl was analyzed without matrix separation by high capacity anion-exchange column. Detection limit for Cl was 0.01 µg mL?1 in sample solution and 1 µg g?1 in solid sample. The reproducibility (100 µL injected) was better than 3%, 3% and 4% at 50, 25 and 50 µg L?1 level for Cl, Na, and Mg respectively. The overall precision was better than ±7% for Na-Mg and ±5% for Cl.
Keywords:Chloride in matrix nitrate  high capacity column  ion chromatography  Na and Mg determination  neutron poison  solid phase separation
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号