首页 | 本学科首页   官方微博 | 高级检索  
     


Development of Polymeric Membrane Sensor for the Determination of Histamine: Experimental and Theoretical Study
Abstract:Abstract

Some original histamine membrane sensors were prepared by incorporating thiopyrilium (TP) derivative as an excellent charged carrier. Among these sensors, TP2 shows the best response to histamine in a wide concentration range from 2.5×10?6 to 1.0×10?1 mol L?1 with a slope of 55.6±0.7 mV decade?1 and with a detection limit of 1.0 µmol L?1 (~0.1 ppm). The interference between histamine and several amino acids and common inorganic anions was negligible as shown by the potentiometric selectivity coefficient data. The prepared electrode was used for the determination of histamine in a human serum synthetic sample. Theoretical calculations were carried out to find the best interactions between histamine and different ligands.
Keywords:Sensor  selective  membrane  histamine  potentiometric  thiopyrilium  ab initio calculation
设为首页 | 免责声明 | 关于勤云 | 加入收藏