首页 | 本学科首页   官方微博 | 高级检索  
     


Detection of Anabolic Steroids By Gc/Sim/Ms With Trifluoroacetylation in Equine Plasma and Urine
Abstract:ABSTRACT

A method to detect three anabolic steroids (boldenone, nandrolone and mesterolone) is presented. The anabolic steroids are isolated from equine plasma and urine by extraction with diethyl ether and C18 Sep-Pak cartridge adsorption, respectively. The extracts obtained were derivatized with trifluoroacetyl anhydride and analyzed by GC/SIM/MS. The selected ion monitoring (SIM) mode was applied to increase the sensitivity and, when possible, the higher m/z ions were selected to improve identification. Stability of derivatives was good and compounds having hydroxy and conjugated ketone groups produced trifluoroacetyl ester derivatives that were also stable. Repeatability of the chromatographic analysis was evaluated on the basis of area repeatability, and the coefficient of variation obtained was lower than 4.4. The detection limit was 1 and 5 ng/ml for all the anabolic steroids studied in equine plasma and urine, respectively.
Keywords:Anabolic Steroids  GC/MS  Trifluoroacetylation  Equine