首页 | 本学科首页   官方微博 | 高级检索  
     


Catalytic Determination of Molybdenum by Means of the Sodium Perborate/Bromide/Ascorbic Acid Landolt Reaction Using Potentiometric Measurements
Abstract:Abstract

A new catalytic method for the determination of 0.1-10 ppm molybdenum is described. Molybdenum catalyses the perborate - bromide ascorbic acid Landolt reaction at pH 3.64. The rate is monitored by potentiometry. A calibration graph is based on the ratio of the reaction times for the blank and the sample, (t(o)/t(c), plotted against the concentration of molybdenum. Optimal conditions for the determination are discussed. Also the effect of various ions and ihterferences are described.
Keywords:Landolt reaction  Molybdenum  Sodium perborate  Ascorbic acid  Reaction times  Potentiometric measurements
设为首页 | 免责声明 | 关于勤云 | 加入收藏