Abstract: | Abstract A sensitive and selective spectrophotometric flow injection analysis (FIA) method with chlorophosphonazo-mN has been developed for the determination of uranium(VI) in standard ore samples. Most of interfering ions are effectively eliminated by the masking reagent of diethylenetriaminepentaacetic acid (DTPA). In the U(VI)-chlorophosphonazo-mN system, the maximum absorption wavelength is at 680 nm and Beer's law is obeyed in the range of 1 to 15 μg ml?1. The correlation coefficient of the calibration curve is 0.9998, the sampling frenquency is 60 h?1, and the detection limit for uranium(VI) is 0.5 μg ml?1. The composition of the U(VI)-chlorophosphonazo-nN complex was established to be 1:2 by flow-through spectrophotometric and conventional molar ratios methods. |