Rapid and Highly Sensitive HPLC and TLC Methods for Quantitation of Amlodipine Besilate and Valsartan in Bulk Powder and in Pharmaceutical Dosage Forms and in Human Plasma
Abstract:Two simple, sensitive, and specific high-performance liquid chromatography and thin-layer chromatography methods were developed for the simultaneous estimation of Amlodipine besilate (AM) and Valsartan (VL). Separation by HPLC was achieved using a xTerra C18 column and methanol /acetonitrile /water/ 0.05% triethylamine in a ratio 40:20:30:10 by volume as mobile phase, pH was adjusted to 3 ± 0.1 with o-phosphoric acid. The flow rate was 1.2 mL min?1. The linearity range was 0.2 to 2 µg mL?1 for amlodipine besilate and 0.4 to 4 µg mL?1 for Valsartan with a mean percentage recovery of 99.59 ± 0.523% and 100.61 ± 0.400% for amlodipine besilate and valsartan, respectively. The TLC method used silica gel 60 F254 plates; the optimized mobile phase was ethyl acetate/ methanol / ammonium hydroxide (55:45:5 by volume). Quantitatively, the spots were scanned densitometrically at 237 nm. The range was 0.5–4.0 µg spot?1 for amlodipine besilate and 2.0–12.0 µg spot?1 for valsartan. The mean percentages recovery was 99.80 ± 0.451% and 100.61 ± 0.363% for amlodipine besilate and valsartan, respectively. The HPLC method was found to be simple, selective, precise, and reproducible for the estimation of both drugs from spiked human plasma.
Keywords:Amlodipine besilate  HPLC  TLC  Valsartan