首页 | 本学科首页   官方微博 | 高级检索  
     


Response Amplification of an Ion-Selective Electrode
Abstract:Abstract

The response of an ion-selective electrode can be amplified by connecting the cell, which is composed of an ion-selective electrode and a reference electrode, in series. For a cell using 1, 2 or 3 valinomycin electrodes connected in series, the response slopes to 1 × 10?5 ?1 × 10?1 M K+ were 58, 116 and 174 mV/activity decade (a.d.) at 25°C, respectively. This amplification method would especially be useful for accurate determinations with electrodes in the range of low concentrations outside the Nernstian response or for determinations of polyvalent ions, in which both cases exhibit small emf response changes in changing the ion concentration.
Keywords:Amplification  ion-selective electrode  response  valinomycin electrode